Fat Suppression Techniques
Greg Brown vision@adelaide.on.net
 
Introduction
Principles  
Relaxation Rates 
Chemical Shift 
Spectra Display Convention
Phase Cycling 
Bright fat in Turbo Spin Echo sequences 
J-Coupling
Fat Suppression Techniques  
Fat Saturation  
Fatsat & Field Strength 
Silicon Suppression 
Fatsat - Key points 
Problems with Fatsat 
Field Shimming 
Patient shimming 
Shimming the MAGNETOM 
Intrinsic field inhomogeneity 
Saturation bags 
RF Inhomogeneity
Selective Excitation Routines   
Phase Cycling Gradient Recalled Echo   
STIR Short TI Inversion recovery  
Selecting TI 
TI and Turbo Spin Echo 
STIR and Gadolinium 
STIR - Key Points 
Spectrally Selective Inversion  
Subtraction
MTC and Fat Signal Conclusion
Spin Echo as a Fat Suppression Method References
  Back to Greg's MRI Home Page

Introduction
 
  Fat Path
T1 TSE ++ --
PD TSE ++ +/-
T2 TSE ++ ++
T1 SE ++ --
PD SE + +/-
T2 SE +/-  
STIR -- ++
MP-RAGE ++ -
T1 FLASH + -
FLAIR ++ ++
T1 Gd  ++ ++
Signal from fat plays a critical role in determining the useful contrast of images in many situations. Being able to control the signal developed by fatty tissue is a powerful and possibly under-rated method in the MR imaging repertoire. These notes describe various fat suppression techniques, important characteristics of the MR behaviour of fat and water based hydrogen nuclei, and discuss how these techniques can be used creatively.

Fat signal is visible on most MRI images.

· fat in pathology
· pathology in fat
· fat outlining anatomy and pathology
· fat mimicking pathology
· fat obscuring contrast enhancement
· fat related artefacts
The signal of fat can be helpful in the display of anatomy and pathology. For example in T2 TSE images of the lumbar spine, where bright fat outlines the exiting nerve roots in a manner normally associated with T1 images and the high signal CSF provides high contrast to display the intra-thecal nerve roots clearly. Low contrast between fat and pathology is also encountered and can affect the utility of an image in many clinical settings. Low fat/water contrast is most commonly encountered in T2 weighted TSE sequences, but can also present difficulties in T1 weighted images after paramagnetic contrast administration of when detecting subtle lesions in the marrow space or breast.
In these situations, fat suppression techniques add a useful extra dimension to the manipulation of image contrast. The methods, like much of MRI, have been developed from practice in magnetic resonance spectroscopy. Their general implementation in MRI has depended on hardware developments particularly in magnetic field homogeneity, gradient control, and RF pulse shape control.
TOP

Principles

Fat suppression techniques work by taking advantage of two characteristic differences in the MR behaviour of fat and water, Relaxation rates (T1 & T2), and Chemical Shift effects
 
  T1 T2
Fat 250 60
White Matter 700 75
Tumour 1000 100


Relaxation Rates

Fat has a very short T1 relaxation rate, thus it returns a high signal on most T1 weighted sequences. Its T2 value, as observed in SE images, is slightly shorter than that of most tissues, so fat will display moderate to low signal intensity. In sequences that apply multiple 180O pulses rapidly (TSE, FSE), the observed T2 times of fat change significantly .


Chemical Shift

The MR signals from fat and water have slightly different frequencies. This difference is called the chemical shift. The chemical shift effect results from differences in the distribution of the electron cloud surrounding Hydrogen nuclei in the molecules of` water and fat. The external magnetic field causes an additional drift in the electron clouds of molecules. This drift of charged particles creates a magnetic field that opposes the external field (Lenz's Law) and thus decreases the strength of the magnetic field experienced by nearby nuclei. The electrons are said to "shield" the nucleus. The strength of this opposing field is proportional to the strength of the external field and the density of the electron clouds surrounding the nuclei. A stronger shielding field reduces the experienced magnetic field so the nucleus will precess at a lower Larmor frequency.

Fat/Water Chemical Shift
 
Field Strength Freq. difference
1.5 T 223 Hz
1.0 T  150 Hz
0.5 T 75 Hz
In fat, the hydrogen nuclei are located in a -CH3 radical attached to a large organic molecule. The electron clouds are shared quite evenly and the Hydrogen nuclei experience some shielding. In the water molecule (H2O) the strongly electro-negative Oxygen atom "steals" the electrons from the Hydrogen nuclei leaving them relatively de-shielded and therefore experiencing a higher magnetic field that the hydrogen in fat.

Chemical shift is expressed in parts per million (ppm), the absolute frequency difference is proportional to the main magnetic field strength. Water precesses approximately 3.5 ppm faster than fat.
TOP


Spectra Display Convention
Figure 1  Spectra Display

By convention MR spectra (figure 1) are displayed with a reversed frequency axis. The higher frequency peaks are plotted to the left of the page. In many texts the sign of the axis is left out completely leaving you to assume incorrectly that fat has a higher resonant frequency than water. The confusion is further compounded by texts that only discuss the difference in frequency and not the mechanism for it.
The convention results from the methods used to obtain most early NMR spectra, in which the RF system operated at a fixed frequency and the magnetic field was swept from low to high values. Thus species with a higher apparent Gyromagnetic ratio would resonate at the fixed frequency with a lower applied external field, appearing earlier in the field sweep and therefore on the left of the display.(1)
TOP



Phase Cycling
When fat and water based nuclei are within the same voxel, the differences in their precessional frequencies will cause the MR signals from each to move in and out of phase with each other. This constructive and destructive interference causes the signal level to vary in strength cyclically after the excitation pulse. The frequency of this variation is equal to the chemical shift and so varies with field strength.
The phase cycling effect is exploited in a number of fat suppression routines, notably opposed phase GRE, Dixon methods and Binomial excitation.
 
FIELD In Out In Out In Out In
1.5 T 0 2.2 4.4 6.8 9 11.2 13.4
1.0 T 0 3.4 6.8 10.2 13.6 17 20.4
0.5T 0 6.8 13.6 20.4 27.2 34 40.8
Figure 2 Phase Cycling
TOP

Bright fat in Turbo Spin Echo sequences

There are two reasons why fat appears bright in virtually all sequences that rapidly apply multiple RF pulses; Magnetization Transfer (MT) effects and J-Coupling. The multiple RF pulses act as off-resonance MT pulses, saturating the bound pool of protons. MT induced signal loss occurs in most stationary tissues but not much in fat, leaving it relatively brighter. The more significant effect is based on a phenomenon know as J-coupling or Scalar Coupling.

J-Coupling

J-Coupling or Scalar Coupling is the most significant reason why fat appears bright on RARE (Turbo) sequences, regardless of their contrast weightings. Through interaction with the electron cloud, a nucleus can alter the magnetic field strength experienced by another nucleus on the same molecule. Therefore various nuclei on the same molecule will precess at slightly different frequencies, and will interfere constructively and destructively in a cyclic fashion with each other (beat). In fat molecules, this results in some T2 shortening, which causes a decrease of signal intensity in images obtained with TE values between 25-100 msec. In Turbo sequences refocussing pulses are typically applied with an echo spacing of less than 25 msec. This repeated refocussing neutralises the J-coupling mechanism of T2 signal loss, and so the fat signal can increase by about 50% compared to SE images. The J-coupling effect is independent of field strength, so the increase of fat signal is observed equally when using TSE sequences on all systems.
TOP



 
Chemical Shift T1 times Miscellaneous
Fat Saturation STIR Subtraction
Selective Excitation SPIR Spin Echo
Phase Cycling GRE   MTC
Dixon Methods    
RODEO 
SPIR
   
Fat Suppression Techniques

Commercially available fat suppression techniques are known by a number of vendor specific acronyms. This section of the paper will group them by the underlying principle used and try to clear the confusion. Not all of these methods are discussed in this paper.
 
 
 
 

TOP


Fat Saturation
Also known as: FATSAT (FS), CHEMSAT, CHESS.

Figure 3  Fatsat GRE Pulse DiagramThese techniques are known generically as Spectral Saturation routines. The method takes advantage of the difference in resonant frequencies between water and fat. A 900 RF pulse, tuned to the resonant frequency of fat is applied flipping the bulk magnetic vector from fat into the transverse plane. Spoiler gradients are then applied to destroy the phase coherence of the signal. This saturation routine is followed immediately by the imaging sequence and the images will only show signal from the remaining water nuclei. Fat based nuclei will not produce a signal until there has been time for significant T1 based recovery (approx. 100 msec).
By adjusting the frequency and bandwidth of the saturation pulse water or silicon can be selectively saturated.
The great attraction of the Fatsat technique is that it only modifies the signal of fat, all other contrast relationships and signal characteristics remain the same. Fatsat can be applied to virtually any sequence with any weighting.
Fatsat is simpler and faster to accomplish at higher field strengths because the separation of the fat and water peaks is greater in absolute terms, and wider bandwidth or shorter period pulses can be used.
The application of saturation pulse and gradients takes a considerable time, reducing the time available for acquiring multiple echoes or slice locations within a given TR. On the Siemens VISION scanner the Fatsat routine requires 19 msec per slice in conventional sequences, and 10 msec per slice for Turbo sequences. In many cases TR must be increased or slices sacrificed to implement Fatsat which can cause problems particularly with short TR techniques. Sequence designers usually apply a Fatsat routine preceding each excitation for each slice but when time savings are important they may opt to apply them less frequently.
Fatsat increases the SAR value of a sequence significantly because it must apply an extra 900 pulse for each slice every TR period.
TOP



Fatsat & Field Strength

The Fatsat routine is more difficult to apply at lower field strengths simply because the fat and water resonant frequencies are closer together. The time required to run the spatial saturation routine is longer at lower filed strengths for two reasons. To achieve the same bandwidth the RF pulse period must be increased by the inverse of the change in field strength, thus a 10 msec pulse at 1.5T will need to be replaced by a 30 msec pulse at 0.5T 2. The pulse period must be further increased if the bandwidth is reduced to accommodate the small absolute chemical shift. Sequence designers must deal with these conflicts, but it is one reason why Fatsat is not commonly employed for T1 imaging on lower field strength machines.TOP



Silicon Suppression
The spectral saturation routine can be tuned to silicon or water frequencies if required, but this is technically more difficult than Fatsat as silicon has a chemical shift of only 4.8 ppm which is only 1.3 ppm more than the fat signal (80 Hz at 1.5T). The silicon suppression technique has been applied commercially to assist in diagnosis of breast implant rupture. In the Siemens breast imaging sequences (tirm11_ss_60b130.wkc) Fat signal is suppressed by a STIR routine and Silicon is suppressed by a spectral saturation routine.3



Fatsat - Key Points · Only modifies the appearance of fat
· Works well with almost all acquisition sequences
· Requires a very uniform magnetic field
· Best at high field
· Increased SAR and scan times
· Works after Gd enhancement
TOP

Problems with Fatsat

Figure 4 Fatsat & field varaitionsThe saturation pulse is applied to the entire imaging volume but it will only work if the frequencies of the pulse coincide with the resonant frequency of fat. Uniform saturation of fat signal all over the image and through each slice depends on achieving an excellent magnetic field uniformity. By convention the bandwidth of the Fatsat pulse is equal to the chemical shift between water and fat, so field variations of more than 1.75 ppm will degrade the Fatsat. It is also possible to get accidental saturation of the water signal where the field strength varies positively. Accidental water saturation is less likely if a narrow Fatsat pulse is used, but this requires a longer pulse period and will result in a loss of fat saturation at even lower levels of field variation.
TOP



Field Shimming
Base magnet shimming provides the foundation to reliable robust Fatsat. During installation the magnet will be shimmed to compensate for intrinsic inhomogeneity and for the effect of the surrounding environment, a process that often stops when the shim figures fall within current acceptance limits. Frequently better shim can be achieved with one or two extra iterations of the procedure.
Moving vehicles and changing external environments can all alter the base shim of the magnet particularly for unshielded and simple self shielded magnets. There is also evidence that magnets slowly alter their shim characteristics making long term monitoring of base magnet shim advisable. These problems can eventually get to the point where the active shim systems cannot adequately compensate and a new passive shimming is required.
Systems installed before the availability of Fatsat sequences may not be shimmed to the same specifications as the current models. The homogeneity figures of such systems should be checked against any revision of base acceptance criteria.
You should regularly check the bore and outside of the magnet for stray metal (paperclips etc), and additionally if you suddenly experience Fatsat problems.
TOP


Patient Shimming
Patient shimming is essential whenever Fatsat is used. Ferro-magnetic material within the scan room can usually be compensated for, but metal in patients' clothing, attachments, or in pillows and pads will affect Fatsat performance. Patient shimming generally uses offset currents in the gradient coils to compensate for linear field variations. In extreme situations this approach won't correct all inhomogeneity or the required level of offset currents will begin to limit the capacity of the gradient systems. In these situations more complex shim routines, altering the current in the higher order shim coils may be necessary. These procedures vary for each machine and require a good understanding of the process to avoid problems. In general it is more appropriate for most users to employ another form of fat suppression.
TOP


Shimming the Siemens MAGNETOM
Sequences stored with the Shim Adj. box ticked will perform a Global or MAP-shim if this has not been done previously. The automatic MAP-shim will only perform a specific number of attempts (iterations) to achieve a well shimmed field. If these attempts are unsuccessful you will see the message "Gradients could not reach convergence". In this situation, or whenever the region is likely to contain extreme intrinsic filed inhomogeneity (breast, neck) it is advisable to perform a MAP-Shim manually through the Adjustment function. Read sections B.14 of the Numaris Reference Guide before doing this. Simply keep selecting the Start button until all the selected channels (typically Gx,Gy,Gz) show a # mark, then press Apply and Exit. The trick is to be persistent, Achieving convergence in breast imaging can require up to 20 iterations but its essential if you are using spectral based fat suppression methods.

VISION users can choose to include shim coils in the shimming process. The software will then attempt to optimise the shim of the local Field of View in different directions depending on which shim coils are selected. This adds to the quality of the shim as well as to the time needed to get convergence. With software B31A select only from channels A20 A21 B21 B22.(4) If the A20 (Z2) channel is altered you must repeat the frequency adjustment process manually before starting the sequence.

Shimming difficult regions is often unsuccsesful when using an array coil or surface coil, particularly if there is fat in the near region of the coil, or the region of interest is displaced from the iso-centre. In these cases the body coil should be used for the shimming process. This technique is more fully described in a notes elsewhere on this web site (Getting Better Fatsat) Changes to the MAP-shim software are expected in NUMARIS B33A.
TOP



Intrinsic field inhomogeneity
Most regional failures of Fatsat occur because of magnetic susceptibility interfaces in the patient. Where bone or air meets
tissue, the sudden change in magnetic susceptibility causes gross regional field gradients and a loss of fat suppression showing as unexpected high signal. These are generally predictable and can be "reported through". If the artefact may be clinically significant another fat suppression technique needs to be used. Typically fat suppression may be lost adjacent to the paranasal sinuses, the lower neck and supra-clavicular region, breasts, inner thighs, and around any implanted metal.


 Fatsat Inhomogeneity Sources

Saturation bags
Saturation bags are designed to remove external air/tissue interfaces and can help to maintain uniform fat saturation around the neck, buttocks and limbs. Users report various degrees of success with a range of filling material. The bags usually contain liquid suspensions of Kaolin (Kaopectate), but fine powdered kaolin available commercially as "china clay" is also effective if sieved and filtered for ferro-magnetic contaminants. The main difficulty of saturation bags is keeping the liquid or powder material contained inside the bags.
TOP


RF Inhomogeneity
Errors can also result from poor distribution of the RF energy (B1 inhomogeneity). These should be corrected by the sequence designer but might be seen with custom or experimental sequences and coils. Surface and array coil images will often show loss of Fatsat in the near regions of the coils. These may result from distortion of the B1 field by the coil or excessive sensitivity of the coil in its near region, at this stage the author has no conclusive information, but recognising the effect as a known artefact can avoid confusion.


Selective Excitation Routines

As was discussed earlier, a frequency selective RF pulse needs a longer time period than a broadband pulse. The binomial excitation technique uses a series of correctly phased broadband RF pulses to selectively excite water by exploiting the phase cycling effect. They achieve selective suppression in a shorter time period than spectral suppression and with lower SAR.

Binomial excitation methods are described numerically with the number referring to the relative amplitude of the pulse. A bar printed above the number may be used to indicate a phase reversal.
Figure 5 Binomial excitationThe simplest is 1:1 excitation also known as jump back excitation.  In 1:1 excitation two 450 broadband pulses are applied with an inter-pulse delay (t) equal to the time taken for fat and water to get out of phase. The first pulse is applied along X axis flipping fat and water vectors 450 and aligning their transverse components with the X axis. During the inter-pulse delay the two vectors get progressively further out of phase until the fat vector is lying along the -X axis and the water vector is again along the +X axis. At this point the second pulse is applied with opposite phase to the first (-X), pushing the fat vector back to Mz (flip = 00), while the water vector is pushed to a flip of 900.
The appropriate inter-pulse delay will depend on the field strength and the MR species being considered so it is frequently expressed as the relative phase difference between the vectors expressed in radians. The time needed to achieve 1800 phase separation would be labelled . In essence the binomial excitation routine pulses are phase locked to the target species, with each successive pulse counteracting the previous pulse for that species and adding to the flip angle of the non-target or off-resonance species. In MRI binomial pulse trains are currently used to selectively suppress fat for 3D sequences 5, in spectroscopy they are typically used to eliminate the water peak. The approach can be sped up by decreasing the inter-pulse delay to/2 and applying the second pulse along the -Y axis (6), work is also aimed at applying spatially selective binomial excitation routines to replace spectral saturation . Higher order binomial excitation routines (1:2:1 or 1:3:3:1) achieve a more selective suppression but take more time because of the additional inter-pulse delays.

Binomial Excitation - Key points

TOP

Phase Cycling Gradient Recalled Echo

The phase cycling effect is commonly exploited with T1 weighted spoiled gradient echo sequences (SGE, FLASH, SPGR) when imaging the abdomen and marrow spaces. In Phase (IP) or Opposed Phase (OOP) images can be selected by choosing an appropriate TE. Images obtained when the fat and water vectors are in-phase will have similar contrast to SE images. When the TE coincides with an opposed phase period a distinct dark line (Indian ink artefact) is noted at borders of fatty and water based tissue. This occurs where the voxel contains equal amounts of fat and water spins, achieving a complete cancellation of signal. In voxels where the mix is not 50:50 varying degrees of signal loss will occur compared to the IP image. This signal loss can be used to identify fatty infiltration of the liver, and tumour infiltration of marrow spaces.

T2* decay causes a rapid decrease in Mxy as TE extends in GRE sequences, so it is important to image quality to use the earliest TE possible for the OOP image. At low field strengths this is easily achieved although the OOP time is late. At 1.5T the first OOP time occurs at 2.2 msec and partial echo techniques are often employed to implement this short TE. While this approach sacrifices some SNR, the loss is less than waiting until the next OOP period and suffering the T2* decay.

TOP


STIR

STIR is an acronym for Short TI Inversion Recovery. STIR sequences use an inversion pulse and time delay (TI) to null longitudinal magnetisation (Mz) so that fat will not contribute signal in any subsequent signal generation routine. STIR is most commonly explained with reference to an inversion recovery spin echo (IRSE) sequence but the approach is also applicable to TSE, GRE, GRASE TurboFLASH and EPI acquisitions.

Figure 7 IR Pulse DiagramThe initial 1800 inversion pulse flips longitudinal magnetisation to the -Mz, after which longitudinal recover occurs with the MZ evolving through Mz=0 back to thermal equilibrium. After a time delay (Inversion Time TI) a 900 pulse is applied converting Mz into detectable Mxy. Contrast between tissue depends entirely on T1 differences at this stage. In order to null a tissue the TI must be chosen to coincide with the T1 curve of the tissue crossing the Mz=0 line. For the situation where TR - TI is long (more than 5 times T1) (7), this can be calculated as : TI null = ln2 x T1 or 0.69 x T1

Figure 8 STIR Magnetization CurvesWhen creating STIR images the reconstruction process is only sensitive to the magnitude of the Mz of a tissue, not whether it is positive or negative. It is assumed that fat will have the shortest T1 in the area of interest so while all other tissues have negative Mz at the TI needed to null fat they are displayed with a pixel brightness proportional to their absolute (modulus) value. This type of display is variously called Modulus (IRM) or absolute (IRABS), and is different to the approach used with IR sequences employing more moderate inversion times. Thus in a STIR image, tissues with short T1 are displayed darker than tissues with long T1, which is the opposite of what is expected in a conventional T1 weighted image. As TE is increased T2 effects increase the contrast between tissues because tissues with short T1 typically have relatively short T2. For this reason STIR sequences are often said to have additive T1 and T2 contrast. Late echo STIR images closely resemble Fatsat T2 weighted images and can be applied interchangeably in many applications. Remember however that overall signal levels still fall as TE is increased.
TOP


Selecting TI
 
Field TI (msec)
1.5 T 140-170
1.0 T 130-160
0.5 T 110-140
There is great variety in suggested TI values. Experimentation is the key to obtaining an appropriate value. If the TI is shorter or longer than the exact TInull fat will return some signal, increasing with the difference between TInull and the TI used. This can be employed to select the degree of fat signal suppression required. Some users prefer to retain a little signal from fat in order to aid in localising anatomy especially on long TE images and musculo-skeletal application. Others prefer to use whatever TI is required to totally null fat signal.
In general the appropriate TI will decrease at lower field strengths, because T1 varies proportionally with B0. TInull is also proportional to TR to a degree, although this dependence stops when TR is long enough to allow complete recovery of Mz for fat (TR-TI>5000 msec) (8)


TI and Turbo Spin Echo
The TI required to null fat (or any tissue) is always shorter when a modified RARE sequence (TurboSE, TurboIR, HASTE) is used. The shortening is more pronounced with longer ETL (higher Turbo factor) and longer TE sequences. This results from the repeated inversions of the recovering Mz by the multiple 180 degree pulses used in the Turbo routine allowing the tissue less time to recover than is predicted by the TR value alone. In essence the tissue can only recover during the period between the last180 degree refocussing pulse and the Inversion pulse, which an be estimated as TR - (TI+ TE last) where TE last is the echo time of the last spin echo in the RARE train. Thus TR, ETL and the echo spacing (ESP) affect the TI(9)

TOP


STIR and Gadolinium
Because the signal nulling is based on the T1 value STIR is not a reliable method for displaying Gd contrast. These areas will suffer severe signal suppression and show poor contrast with fat or a total loss of signal. The same mechanism makes STIR inappropriate for detecting other short T1 lesions such as sub-acute blood or melanin.

STIR - Key Points 
  • Fat suppression is based on T1 behaviour and selection of TI 
  • Reverse T1 contrast plus T2 contrast 
  • Uniform fat suppression independent of magnetic field inhomogeneity 
  • Can be implemented at any field strength with equal success 
  • Works well with many acquisition regimes 
  • Don't use STIR post Gd contrast 
  • Higher SAR than Fatsat or Water Excitation
TOP

Spectral Inversion Recovery

This method is a combination of the spectral saturation and STIR routines. It is available on Philips scanners where it is designated Spectral Inversion Recovery (SPIR)(10), and GE scanners with the designation SPECtral Inversion At Lipids (SPECIAL). (11) The idea is to apply a spectrally selective pulse to flip fat spins then, after the time interval that lets the Mz of fat reach zero, the excitation pulse is applied and the signal of the water spins give most of the signal.

SPECIAL image breastTime is the major disadvantage of this fat suppression method. The routine is best applied at least once per TR, and for multi-slice sequences once per slice per TR. The best fat suppression is achieved with a 180 degree spectral inversion pulse and a time delay equal that used in normal STIR imaging ( approximately 150 msec at 1.5 T plus the finite time required for the inversion pulse) This is clearly impractical for anything but extremely long TR techniques.

In practice the "inversion" pulse used ranges between slightly more than 90 degrees but significantly less than 180 degrees the required delay time can be acceptably short. In most scanners using this method the radiographer has control of the inversion flip angle and the system determines the optimum delay time. The routine is applied in a "segmented "fashion allowing some small variation in the degree of fat suppression across slices.

SPIR will be as sensitive to local field inhomogeneity as spectral saturation routines. It is suitable for use after Gd contrast because only the fat spins are affected by the routine
TOP


Subtraction

Fat suppression by subtraction of the pre-contrast image from the contrast images particularly appropriate in the breast where the anatomy is prone to extreme magnetic field inhomogeneity even after patient shimming, and where the dynamic nature of contrast uptake precludes STIR imaging with sufficient spatial and temporal resolution.
Movement will mar subtraction images so patient comfort, immobilisation and procedure speed are very important. The MR receiver gain setting must not be altered between pre and post contrast sequences so that global signal intensity changes caused by the contrast are not neutralised.
 
Pre-contrast Post Contrast Subtraction

Creating subtraction images from a dynamic string of T1 3DFLASH sequences with a fast high dose(.166mmol/kg) injection of Gd-DTPA allows improved lesion detection and discrimination between malignant and benign breast lesions (12). The approach has also been applied with 1.5 second serial scanning to investigate pelvic malignancies. (13)
TOP


MTC and Fat Signal

Magnetisation Transfer Contrast pulses are frequently used to diminish background signal in angiography sequences and post contrast. Fat is largely unaffected by these pulses and so its signal increases relative to the non-fatty tissue. This can be counterproductive in neck angiography and in detecting enhancing lesion in or near fat. In these instances MTC is not recommended.
TOP


Spin Echo as a Fat Suppression Method

At present the bright fat signal on T2 weighted TSE images is often noted as the most obvious difference to a T2 weighted SE image, but increasingly TSE sequences are becoming the standard method of acquiring T2 weighted images. In that event, Spin Echo becomes is a niche sequence, and the fat signal suppression due to scalar coupling effects and a lack of MTC induced contrast must be considered amongst its characteristics. SE offers mild fat signal suppression in a T2 weighted image with reduced SAR loading rates, increased sensitivity to blood breakdown products and a substantial increase in sequence time.


Conclusion

MRI Radiographers have access to a range of fat suppression methods, which work by exploiting particular characteristics of the MR signals from fat and water. Understanding the mechanisms of each method provides insight into appropriate applications and situations where artefacts can be anticipated. The widespread use of TSE sequences for T2 weighted imaging has increased the situations where high fat signal may obscure lesion visibility. Controlling fat/water contrast can enhance the sensitivity and specificity of an MRI examination in many clinical settings, so the MR Radiographer must be familiar with available fat suppression techniques and know when to apply them.
TOP


Acknowledgments

Paper prepared initially for the first New Zealand MR Users' Meeting, Auckland November 1996, then expanded for the Siemens MR User meeting Sydney April 1998.
 
Click Me to Mail to Me 
 If you would like this paper presented at your MRI meeting or you have comments or suggestions, please contact the author by e-mail. 
TOP


References
1. Personal e-mail with C. Gregory University Illinios, April 12 1996    back to text
2. Magnetic Resonance Imaging. M.T. Vlaardingerbroek, J.A. den Boer. Springer Berlin 1996 p 78 back to text
3. Siemens MAGNETOM Applications Guide VB31B I.5-10 back to text
4. Siemens MAGNETOM Numaris Reference Guide VB31B B.14-5 back to text
5. Siemens water excitation (WE) sequences personal e-mail with P.Finn Siemens, Iselin Nov 96 back to text
6. Thomasson D , Purdy D Moore J Finn P Minimum Time spatial spectral pulses using a phase modulated 1-1 binomial pulse design. In Proceedings of the second annual meeting of the Society of Magnetic Resonance, San Francisco 1994 p120 back to text
7. E-mail correspondence E. Kanal 12 Nov 1997 back to text
8. Michael Kean Experimentation on SIGNA Echospeed Royal Childrens Hospital Melbourne October 1996 back to text
9. E-mail on MRI-Technologists list E.Kanal November 1997 back to text
10. 2.4.1 SPIR -Fat Suppression. Philips Gyroscan NT Application Guide Volume 2 P 66-68 Philips Medical Systems 1994 back to text
11. SPECIAL In Breast Imaging Clinical Studies SIGNA Tip Applications Guides CD-ROM GE Medical Systems 1997 back to text
12. Based on work by S.H. Heywang-Korbrunnner R. Beck Contrast Enhanced MRI of the Breast 2nd Edition Springer Berlin 1995 back to text
13. Barentsz J. Book of Abstracts Innaugral Gold Coast MRI and CT Conference August 1996 back to text