Better
Shim Adjustments & Fat Suppression
© Greg Brown
vision@adelaide.on.net
This note explains how to do get good fat suppression when using the
array or surface coils on the Siemens VISION.
It includes
Back to the Adealide MRI Website
When to Expect Trouble with Fat Suppression
Sequences stored with the Shim Adj. box ticked will perform a Global
shim or MAP-shim automatically if this has not been done before for this
patient. The automatic MAP-shim will only perform a specific number of
attempts (iterations) to achieve a well shimmed field. If these attempts
are unsuccessful you will see the message "Gradients could not reach convergence"
and other action will be needed to get good spectral fat suppression (Fatsat).
Fatsat will fail wherever there is more than 1.75 ppm variation in field
homogeneity.
This commonly occurs
-
about 40 cm from the magnet iso-centre in any direction
-
near significant air tissue interfaces such as around the shoulders, skull
base, nasopharynx and breasts.
Getting Better FATSAT Results with Array or
Surface Coils
Use the body coil to receive when doing the shim adjustment. Remember
the system will only perform one shim adjust. If the system has already
shimmed the patient you must use the manual method.
Automatic Shim with body coil
The simplest way to do this is: -
-
Set up the sequence and switch to the Sequence Change platform
-
Tick the Bodyscout box, select Load
-
Watch the screen to confirm that a shim adjustment is performed successfully
(if not you must do a manual MAP-Shim)
-
Untick the Bodyscout box and confirm the array segments selected.
-
Start the sequence with Measure, not Start
Manual MAP-Shim with Body Coil
A more thorough approach is to do a manual MAP-Shim with the body coil.
-
Set up the sequence and switch to the Sequence Change platform
-
Tick the Bodyscout box
-
Select Adjust. At the Adjustment platform select MAP-Shim
-
Select Start and look at the numeric readout in the upper right
half of the platform. You need a hash mark (#) next to each shim channel
selected (X, Y, Z plus any selected in the lower right of the screen).
Be persistent. It can take 30 - 50 attempts to get shim convergence.
-
The message "Shim Adjusted" will appear in the bottom right line of the
platform
-
Select Apply, then Exit to return to the adjustment platform.
-
Confirm the Centre Frequency (CF) is still adjusted to the water signal
(see Confirming CF), then select Change to return to the Protocol
Change platform.
-
Untick the Bodyscout box and confirm the array segments.
-
Start the sequence with Measure (not Start).
The process is detailed in sections B.14 of the Numaris Reference Guide.
Why does it help?
Because these coils have spatially varying sensitivity they can be swamped
by strong signals from fat near the coil. This gives distorted information
for the MAP-Shim to work with and increases the likelihood of reading the
fat rather than the water signal. The frequency adjustment routine tries
to adjust the centre frequency of the MR system to the resonant frequency
of water. This is done by analysing the frequency curve of a single slice,
and assuming that the water causes the maximum peak amplitude. The method
will be tricked when there is a low ratio of water to fat signal, or where
there is a lot of fat close to a surface coil.
Confirming the Centre Frequency (CF)
The CF is the frequency that the scanner is tuned to. With poor shimming
or when fat is close to the coil during initial frequency adjustments the
system may select the fat peak instead of the water peak, resulting in
poor images when using spectral fat suppression.
The CF must be confirmed whenever the A20 (Z2) channel is used during
manual MAP-Shim, and should be done after any MAP-Shim to ensure good results.
-
From the Adjustment platform, select Frequency
-
Click Start, repeat until the message "Frequency Adjusted" appears
at the bottom right of the platform.
-
Place the CF check tool on the screen lining up the marks with the end
of the frequency scale.
-
Ensure that the Water peak (the left peak) is aligned with the CF mark.
If it isn't, repeat the MAP-Shim and frequency adjustments or use STIR
or WE sequences to reduce fat signal.
-
Select Exit, then Change
Making a Centre Frequency Check tool
The Easy Way
-
Print this page (select page 3 in your printer dialogue) and photocopy
the image below on to X-ray film or an overhead transparency sheet.
The Hard Way
-
Call up the Map-Shim platform
-
Fix a piece of paper to the screen and mark the endpoints of the frequency
scale/
-
Fold the paper to find the point midway between the scale ends.
-
Use that gauge to place the end and centre marks on a small piece of X-ray
film
Remember
-
When the field is well shimmed the water peaks and fat peaks will be separate.
-
The Water peak appears to the left of the Fat peak (higher frequency)
-
The Water peak is usually higher than the Fat peak, but that depends on
how much fat is within the field of view, and on the coil sensitivity pattern.
Further Reading
Fat Suppression Technique Notes. Greg Brown 1997
Fat Suppression Presentation Slides Greg Brown 1997
Numaris Reference Guide B.14
 |
Click Me to Mail to Me
If you would like this paper presented at your MRI meeting or you have
comments or suggestions, please contact the author by e-mail by clicking
on the picture.
|
 |